Skip to content

12.1 Varying the primary current

It is advised to use different primary settings (L1, L2 and L3 lenses), for covering different primary current ranges.

L1Fixed about 55% of Duo Source HV
L2Used to increase/decrease the primary beam intensity (standard beam)
L30 to 40% of Duo Source HV for high beam density if require. Over 65% of Duo Source HV for small beam with high density. (Imaging beam)

O₂⁺ primary ions, Source HV = +15kV:

Within a given range, the primary beam intensity can be varied through:

- Aperture selection: a smaller D0 aperture reduces both the current and the beam size. A smaller D4 aperture reduces mostly current, but not significantly the beam size. Therefore, it is advised to use mostly the D4 and L2 for changing the primary current.

- Using Lens L3:

○ Low values (usual conditions): L3 values remain to 0, can be increase up to 30% for very high-density beams.

○ High values (to be used only when very small spot diameter is required): the primary beam intensity (but also the beam size) increases with L3.

Note: L3 should not be used between 30% and 60% of the source HV, in order to keep a well-focused primary beam.

In general, the following rules apply for changing the primary current:

- For decreasing the primary beam intensity, do by order until obtaining the requested primary current: increase first L2 (if L3 is not set to 0 decrease L3 value), then decrease D4 and finally decrease D0.

- For increasing the primary beam intensity: do by order until obtaining the requested primary current: increase D4 (if D4 is smaller than 750μm), increase then D0 (if D0 is smaller than 750μm), and finally decrease L2 to 50% of the source voltage and increase L3 (by keeping L3< 30% of the source HV).

Note that changing the aperture selection changes the primary current by discrete steps, while the L2 lens provides a continuous variation. Therefore, the aperture selection and L2 value may need to be combined if a precise primary current value is requested.

For very small beam setting (beam diameter < few μ ms), the following conditions must be selected: high source HV (typically ≥ 10kV), small D0 (typically 400 μ m and under) and D4 (typically 100 μ m), combined with high L2 value (typically above 70% of the source HV). The primary current (and beam diameter) is reduced by increasing L2 value.

O₂⁺ primary ions, Source HV = +15kV:

Schematics of the primary ion beam trajectory (Duo source) using Standard conditions (L1, L2): Standard beams.
Schematics of the primary ion beam trajectory (Duo source) using Standard conditions (L1, L2): Standard beams.
Schematics of the primary ion beam trajectory (Duo source) using Standard conditions (L1, L2) + lens L3: Highest beam density.
Schematics of the primary ion beam trajectory (Duo source) using Standard conditions (L1, L2) + lens L3: Highest beam density.